Automated Pressure Calibrator
CPC 6000
Mensor/WIKA Operating Instructions - CPC 6000 53
39 Errorno? <sp>{string}<cr><lf> Returns CPC 6000 error code and text.
40
Filter {Off, Low, Normal, High} Sets the reading filter.
41
Filter? <sp>{string}<cr><lf> Returns the reading filter.
42 Gasdensity {value} Sets the head pressure gas density in lb/ft^3.
43 Gasdensity? <sp>{value}<cr><lf> Returns the head pressure gas density in lb/ft^3.
44 Gastemp {value} Sets the head pressure gas temperature in °F.
45 Gastemp? <sp>{value}<cr><lf> Returns the head pressure gas temperature.
46
Gateway {string e.g. nnn.nnn.nnn.nnn} Sets the Ethernet gateway address.
47 Gateway? <sp>{string}<cr><lf> Returns the Ethernet gateway address.
48
Height {value} Sets the head pressure height in inches.
49
Height? <sp>{value}<cr><lf> Returns the head pressure height.
50
Highspeed {Yes or No} Sets the active channel to high speed mode:
.006% stablewindow and 1 second stabledelay.
51
ID? <sp>MENSOR,600,{ssssss},{
v.vv}<cr><lf>
Returns the instrument identity where {ssssss}
is the serial number, and {v.vv} is the software
version number.
52 IP {string e.g. nnn.nnn.nnn.nnn} Sets the IP address of the instrument.
53 IP? <sp>{string}<cr><lf> Returns the IP address of the instrument.
54 Keylock {Yes or No} YES to lock, or NO to unlock the on-screen
keys.
55 Keylock? <sp>{Yes or No}<cr><lf> Returns current keylock status as YES or NO.
56
List? <sp>Pri,X,X;Sec,X,X;Bar,1<c
r><lf>
Returns list of available turndowns on installed
sensors in the active channel. X will be non-ex-
istent if the turndown is not available.
57 Listcal? <sp>PRI,{sn},{td},{mm/
dd/yyyy},{td},{mm/dd/
yyyy};SEC,{td},{mm/dd/
yyyy},{td},{mm/dd/yyyy}<cr>
<lf>
Returns the serial number of each installed
transducer and calibration dates for each range.
58
Listrange? <sp>PRI,{td},{min},{max},{td},
{min},{max};SEC,{td},{min},{m
ax},{td},{min},{max}<cr><lf>
Returns the minimum and maximum ranges of
all installed sensors for the active channel.