50 WIKA operating instructions ProcessCalibrator, model CPH6000
11069023.04 11/2012 GB/D
GB
ENTER
ENTER
7.3.6 CALIBRATION mode (calibration of a pressure gauge)
7. Commissioning, operation
Deviation
Deviation
Deviation
No. of test item
and No. of test
point
No. of test item
and No. of test
point
No. of test item
and No. of test
point
Generate the specied
setpoint in acc. with the
test item display using the
pressure generator
If the set value = 0 bar make
sure the calibration assembly
is in a pressure-free condi-
tion/vented to atmosphere
(test item must indicate 0 bar;
zero correction if necessary)
and with
<--
record the values of this test
point
Generate the specied
setpoint in acc. with the
test item display using the
pressure generator
Record the measured values
of this test point
(with
back to previous
test point)
2. Test point
x. Test point
1. Test point
(e.g. 0 bar)
2. Test point
(e.g. 1 bar)