6. Commissioning, operation
6.12 MIN/MAX memory
The calibrator has a MIN/MAX function for capturing the minimum and maximum values
of any displayed parameter.
The MIN/MAX function can be set by stepping through the menu options until "MIN/
MAX" is shown on the display above the F1 key. At this point, pressing the F1 key
will scroll the display through the MIN/MAX values that are stored in the "MIN/MAX"
registers. These readings are live so new MIN/MAX values will be recorded while in this
mode.
To clear the "MIN/MAX" registers, simply press the "CLEAR" key. These registers are
also cleared on power-up or when the conguration is changed.