WIKA operating instructions model CPP1200-X
GB
14076968.01 12/2013 GB/D
8. Faults
Faults Causes Measures
Equipment provides
pressure but pressure
decays when valves A
and B are operated.
Incorrect operating procedure
being used.
Ensure that correct operating
procedure is being followed (see
chapter 6.6).
If unable to locate a cause. Return test pump to
DH-Budenberg/WIKA for
investigation.
Equipment provides
pressure but pressure
decays to lower value
then remains steady.
Insucient liquid in pressure
balance.
Check liquid level in reservoir. Fill
reservoir with correct liquid as
necessary (see chapter 6.5.1).
Air in the system Pre-ll component under test with
appropriate liquid. If necessary
re-ll pressure balance with
appropriate liquid.
If unable to locate a cause. Return test pump to
DH-Budenberg/WIKA for
investigation.
Internal damage Return test pump to
DH-Budenberg/WIKA for
investigation.
Incorrect operating procedure
being used.
Ensure that correct operating
procedure is being followed (see
chapter 6.6).
If unable to locate a cause. Return test pump to
DH-Budenberg/WIKA for
investigation.
The spindle pump of
the test pump becomes
very sti to operate
when the test pump is
being used in the range
below 140 bar (2,000 lb/
in²).
Internal damage Return test pump to
DH-Budenberg/WIKA for
investigation.
The spindle pump of
the test pump becomes
very sti to operate
when the test pump is
being used in the range
over 140 bar (2,000 lb/
in²).
Incorrect operating procedure
being used.
Ensure that correct operating
procedure is being followed (see
chapter 6.6).
If unable to locate a cause. Return test pump to
DH-Budenberg/WIKA for
investigation.