Proceed as follows:
1 Edit the mbar value in the menu item "zero" with [OK].
Zero
000.0 %
=
+0000.0 mbar
0000.0 mbar
2 Set the requested mbar value with [+] and [->].
3 Conrm with [+] and move to span adjustment with [->].
The zero adjustment is nished.
Information:
The zero adjustment shifts the value of the span adjustment. The span,
i.e. the dierence between these values, however, remains u nchan-
ged.
Information:
For an adjustment with pressure, simply enter the actual measured
value indicated at the bottom of the display.
If the adjustment ranges are exceeded, the message "Outside
parameter limits" appears. The editing procedure can be aborted with
[ESC] or the displayed limit value can be accepted with [OK].
Proceed as follows:
1 Edit the mbar value in the menu item "span" with [OK].
Span
100.0 %
=
+1000.0 mbar
0000.0 mbar
Information:
The displayed pressure for 100 % corresponds to the nominal
measuring range of the sensor (in the above example 1 bar =
1000 mbar).
2 Set the requested mbar value with [->] and [OK].
3 Conrm with [OK] and move to the menu overview with [ESC].
The span adjustment is nished.
Information:
For an adjustment with pressure, simply enter the actual measured
value indicated at the bottom of the display.
Carrying out zero ad-
justment
Carrying out span ad-
justment