Install per National Code an local codes, as applicable. Installation should be
in accordance with ANSI/ISA RP12.6
2
Associated apparatus (barrier) must be entity - approved by FMRC, used in
an approved configuration and installed in accordance with barrier
manufacturer's instructions. Transmitter Vmax must be greater or equal to
barrier Vt or Voc.
Transmitter Vmax must be greater or equal to barrier It or Isc. Transmitter
Imax must be greater or equal to barrier It or Ioc. Transmitter Pmax must be
greater or equal to barrier Po.
Total cable capacitance plus Ci of transmitter shall not exceed Ca of barrier.
Total cable inductance plus Li of transmitter shall not exceed La of barrier.
Vmax >= Voc or Vt
Imax >= Isc or It
Pi >= Po
Ci + Ccable <= Ca
Li + Lcable <= La
3
The following temperature ranges are allowed depend on temperature code:
for T12.10.008 for T12.30.008
Temperature code ambient Temperature code ambient
temperature range temperature range
T4 -40 °C to 85 °C T4 -20 °C to 70 °C
T5 -40 °C to 75 °C T5 -20 °C to 70 °C
T6 -40 °C to 60 °C T6 -20 °C to 60 °C
4
Thermometers must be entity - approved by FMRC, used in an approved
configuration and installed in accordance with thermometer manufacturer's
instructions. Thermometer Vmax must be greater or equal to transmitter Vsc.
Thermometer Imax must be greater or equal to transmitter Isc.
Thermometer Pi must be greater or equal to transmitter Po.
Total capacitance of thermometer shall not exceed Ca of transmitter.
Total inductance of the thermometer shall not exceed La of transmitter.
5
No revision to drawing without prior FMRC approval.