The lower limit of the measuring range is congurable between
-50 ...+150 °C.
The possible upper limit of the measuring range is dependent on
the respective lower limit of the measuring range. For possible
combinations of lower and upper limits of the measuring range, see
diagram. For clarity, this dependency has been shown in 50 °C steps
in the examples in this diagram. The conguration software checks
the required measuring range and will only accept permissible values.
Intermediate values are congurable; the smallest increment is 0.1 °C.
The thermometers are delivered with a basic conguration (0 ... +150 °C,
downscale) or congured to customer specications within the
congurable limits.
With conguration to customer specications, the measuring range
will be shown clearly on the instrument label. The conguration and
the choice of parameters is described in the menu-driven conguration
software. Changes to the conguration must be noted on the label using
a water-resistant felt-tip pen.
Possible combinations of the lower and upper limits of the
measuring range
The upper limit of the measuring range is dependent on the respective
lower limit of the measuring range. For clarity, this dependency has
been shown in 50 °C steps in the examples in these diagrams. The
conguration software checks the required measuring range and will
only accept permissible values. Intermediate values are congurable; the
smallest increment is 0.1 °C.
7. Conguration model TR30-W