Parameter
Rel.
Index
PA
Description Type Store
Size
byte
RO /
R/W
Min. Max. Default
CAL_POINT_HI_2
80
The high calibration value applied to sensor 2
The value from either a calibrator or a reference equip
-
ment.
Float SRC 4 R/W 10
38
CAL_ACTUAL_HI_2
81
Entering any value will force the device to automatically
measure and store the actual high point value. Must be
entered with the applied CAL_POINT_HI_2 value
Float SRC 4 R/W 10
38
SENSOR_CAL_METHOD_2
82
Enables or disables the last sensor calibration for
sensor 2
103 = Factory trim standard (calibration values disabled)
104 = User trim standard (calibration values enabled)
Un-
signed
8
SRC 1 R/W 103
SENSOR_CAL_LOC_2
83 The last location of the calibrated sensor
OCTET_
STRING
SRC 32 R/W » »
SENSOR_CAL_DATE_2
84
The last date on which the calibration was performed
7
*
Un-
signed
8
SRC 7 R/W
0,0,0,0,
1,1,103
SENSOR_CAL_WHO_2
85
The name of the person responsible for the last sensor
calibration
OCTET_
STRING
SRC 32 R/W » »