48 WIKA operating instructions ProcessCalibrator, model CPH6000
11069023.04 11/2012 GB/D
GB
optional delay-time [sec]
Input via numeric input keypad and conr-
mation with
ENTER
With the entry of such a delay-time, the acceptance/recording of the test point is
blocked for this duration. (In the above example, after the rst test point has been
recorded, 30 seconds must pass before the second test point can be recorded.)
Should all test points need to be cleared or reset, since the new
calibration consists of fewer test points than the previous test series,
the CLEAR button simply needs to be pressed. This will clear and
reset the current and all subsequent test points. (This process can take
several seconds.)
Should all recorded calibration data for all calibrations need to be
cleared at the same time see chapter 7.4.7 "SETUP additional menu
item: CLEAR CalData".
7. Commissioning, operation