40 WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.02 05/2014 GB/D
GB
13. Heating and cooling times
13.5 Dry-well temperature calibrator, model CTD9100-650
Measuring conditions
All times refer to a Pt100 reference sensor with Ø 6 mm
The reference sensor is positioned, at full depth, centrally
in the insert
All times are transition times and do not account for the
required settling time
The measurements were carried out at a room tempera-
ture of approx. 23 °C
Insert
Heating up: Time Cooling: Time
25 °C to 40 °C 0:54 min 650 °C to 600 °C 2:25 min
40 °C to 50 °C 0:22 min 600 °C to 550 °C 2:33 min
50 °C to 100 °C 1:18 min 550 °C to 500 °C 2:55 min
100 °C to 150 °C 1:06 min 500 °C to 450 °C 3:27 min
150 °C to 200 °C 1:03 min 450 °C to 400 °C 4:01 min
200 °C to 250 °C 1:05 min 400 °C to 350 °C 4:39 min
250 °C to 300 °C 1:06 min 350 °C to 300 °C 5:36 min
300 °C to 350 °C 1:09 min 300 °C to 250 °C 6:46 min
350 °C to 400 °C 1:21 min 250 °C to 200 °C 8:32 min
400 °C to 450 °C 1:30 min 200 °C to 150 °C 11:22 min
450 °C to 500 °C 1:32 min 150 °C to 100 °C 17:01 min
500 °C to 550 °C 1:38 min 100 °C to 50 °C 52:37 min
550 °C to 600 °C 1:55 min 50 °C to 40 °C 15:23 min
600 °C to 650 °C 2:33 min 40 °C to 25 °C 1:01:58 min