8 WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.02 05/2014 GB/D
GB
3. Specications
3. Specications
3.1 Dry-well temperature calibrator, CTD9100 series
CTD9100-COOL CTD9100-165
Temperature range -55 ... +200 °C -35 ... +165 °C
Accuracy 0.15 ... 0.3 K 0.15 ... 0.25 K
Stability ±0.05 K ±0.05 K
Display resolution 0.01 to 100 °C, then 0.1 0.01 to 100 °C, then 0.1
Gradients, axially
1)
< 0.04 K at 200 °C < 0.04 K up to 100 °C
0.06 K up to 165 °C
Heating time incl. stabilisation
2)
15 min from 20 to 200 °C 12 min from 20 to 165 °C
Cooling time incl. stabilisation
2)
10 min from +20 to -20 °C 7 min from +20 to -20 °C
Immersion depth 150 mm 150 mm
Insert dimensions Ø 28 x 150 mm Ø 28 x 150 mm or Ø 60 x 150 mm
Voltage supply
Power supply AC 100 ... 240 V, 50/60 Hz AC 100 ... 240 V, 50/60 Hz
Power consumption 555 VA 375 VA
Power cord for Europe, AC 230 V for Europe, AC 230 V
Communication
Interface RS-485 RS-485
Case
Dimensions 215 x 305 x 425 mm (W x H x D) 215 x 305 x 425 mm (W x H x D)
Weight 11 kg 11 kg
CTD9100-450 CTD9100-650
Temperature range t
amb.
... 450 °C t
amb.
... 650 °C
Accuracy 0.3 ... 0.5 K 0.3 ... 0.8 K
Stability ±0.05 K at 100 °C
±0.1 K at 450 °C
±0.05 K at 100 °C
±0.1 K at 600 °C
Display resolution 0.01 to 100 °C, then 0.1 0.01 to 100 °C, then 0.1
Gradients, axially
1)
0.05 K at 100 °C
0.2 K at 450 °C
< 0.2 K at 100 °C
0.5 K at 600 °C
Heating time incl. stabilisation
2)
14 min from 20 to 450 °C 28 min from 20 to 600 °C
Cooling time incl. stabilisation
2)
58 min from 450 to 100 °C 60 min from 600 to 100 °C
Immersion depth 150 mm 150 mm
Insert dimensions Ø 60 x 150 mm Ø 28 x 150 mm
Voltage supply
Power supply AC 230/240 V, 50/60 Hz AC 230/240 V, 50/60 Hz
3)
(AC 100 ... 240 V, 50/60 Hz)
4)
Power consumption 2,000 VA 1,000 VA
Power cord for Europe, AC 230 V for Europe, AC 230 V
Communication
Interface RS-485 RS-485
Case
Dimensions 150 x 270 x 400 mm (W x H x D) 150 x 270 x 400 mm (W x H x D)
Weight 7.5 kg 8 kg
1) The gradient is understood to be the temperature change in the test well over the rst 40 mm from the bottom of the sleeve.
2) The reference thermometer with which the measurements are performed, has a diameter of 6 mm.
3) Instrument design available with multi-voltage power supply
4) AC 115 V power supply must be specied on the order, otherwise an AC 230 V one will be delivered.