The ventilation of the measuring cell is realised by a lter element in
the socket of the electronics housing. The ventilation of the electronics
housing is realised via an additional lter element around the cable
glands.
2)
1
2
Fig. 3: Position of the lter elements
1 Filter element for ventilation of the measuring cell
2 Filter element for ventilation of the electronics housing
Information:
Make sure that the lter elements are always free of buildup during
operation. A pressure washer must not be used for cleaning.
Higher process temperatures mean often also higher ambient
temperatures.
1
2
Fig. 4: Temperature ranges
1 Process temperature
2 Ambient temperature
Make sure that the upper temperature limits for the environment of
electronics housing and connection cable specied in chapter
"Technical data" are not exceeded.
2)
With previous instrument versions, ventilation and pressure compensation
were carried out together via a lter element.
Ventilation and pressure
compensation
Temperature limits
Process pressure transmitter IPT-1* Vers. 3.0 - metalic sensor 4 20 mA/HART 11
4 Mounting
31543-EN-090130