29WIKA operating instructions, models CTH6300 and CTH6500
2079988.02 04/2013 GB/D
GB
6. Commissioning, operation
Use the arrow keys ▲▼ now to choose between rH for humidity calibration and P °C for
temperature calibration.
Use the arrow keys ▲▼ to select the required calibration function.
Standard characteristic curve oFF
Select oFF by means of the arrow keys ▲▼. Press ENTER/MENU to conrm the
required setting.
Press ESC to return to the measuring mode.
Calibration by code oP1
Select oP1 by means of the arrow keys ▲▼. Press ENTER/MENU to conrm the
required setting. A small 1 appears in the lower part of the display. Four characters
follow it (hex code / 0 ... F). Change the 4 characters by means of the arrow key ▲. Press
the arrow key ▼ to go to the next position. If all 4 characters are entered as required,
conrm by pressing ENTER/MENU. A small 2 appears and the following 4 characters
can be changed as well.
Press ESC to return to the measuring mode.
MENU
Unit Prob Lin2 CAL Chnl ArEA Lo6
CAL
rH P°C
CAL
oFF oP1 oP2