9. Faults / 10. Dismounting, return and disposal
Indication Cause Measures
E1oh
Interruption of the Auto-o
function
Switch on the Auto-o function
again, see chapter 6.7 "AUTO-
OFF function".
E1eh
Interruption of the Auto-o
function
Switch on the Auto-o function
again, see chapter 6.7 "AUTO-
OFF function".
E23
EE-prom content destroyed Send in for repair
E25
EE-prom content destroyed Send in for repair
E12
Overow Remove battery and insert it
again
E31
Temperature of the reference
measuring point is outside of the
measuring range
Use the temperature probe
according to the technical speci-
cations. See chapter 3 "Speci-
cations".
CAUTION!
If faults cannot be eliminated by means of the measures listed above, the
hand-held thermometer must be shut down immediately.
In this case, contact the manufacturer.
If a return is needed, please follow the instructions given in chapter
10.2 "Return".
10. Dismounting, return and disposal
WARNING!
Residual media at the dismounted hand-held thermometer and/or
temperature probe can result in a risk to persons, the environment and
equipment.
Take sucient precautionary measures.