12 WIKA operating instructions precision pressure transmitter model P-3x
11473500.04 10/2011 GB/D/F/E
GB
3.5 Reference conditions (IEC 61298-1)
Temperature: 15 ... 25 °C
Atmospheric pressure: 950 ... 1050 mbar
Humidity: 45 ... 75 % relative
Nominal position: Process connection lower mount (LM)
Power supply: DC 24 V
Load: see output signals
Mounting position: as required
3.6 Operating conditions
Mechanics
Vibration resistance:
10 g (IEC 60068-2-6, under resonance)
Shock resistance:
200 g (IEC 60068-2-27, mechanical)
Service life: 10 million load cycles
Free fall: 1 m
Temperatures
Compensated temperature range: -20 ... +80 °C
Medium: -20 ... +105 °C
Ambient: -20 ... +80 °C
Storage: -40 ... +85 °C
3.7 Electrical safety
Short-circuit resistance: S
+
vs. 0V
CAN-High / CAN-Low vs. U+ / 0V
Reverse polarity protection: U
+
vs. 0V
Overvoltage protection: DC 36 V
Insulation voltage: DC 500 V
3. Specications