22 WIKA operating instructions precision pressure transmitter model P-3x
11473500.04 10/2011 GB/D/F/E
GB
Faults Causes Measures
Constant output signal
upon change in pressure
Mechanical overload caused by
overpressure
Replace instrument; if it fails repeatedly,
contact the manufacturer
No communication Adjustment of node ID
Transmission rate incorrect
Check the conguration parameters
No cyclic transmission PDO mapping
Transmission type incorrect
Check the conguration parameters
Signal span too small/drops Mechanical overload caused by
overpressure
Diaphragm damaged, e.g. due to
impacts, abrasive/aggressive medium;
corrosion at diaphragm/process
connection; transmission medium
missing
Sealing/sealing face damaged/soiled,
sealing does not have a tight t,
threads jammed
Replace instrument; if it fails repeatedly,
contact the manufacturer
Contact manufacturer and replace
instrument
Clean the sealing/sealing face, replace
sealing if applicable
Signal span varies/inaccurate EMC interference sources in the envi-
ronment; e.g. frequency converter
Operating temperatures too high/low
Instrument not earthed
Strongly varying pressure of the
process medium
Shield instrument; cable shield; remove
source of interference
Observe the permissible temperatures
Earth the instrument
Damping; consulting by the manufacturer
Deviating zero point signal Operating temperature too high/low
Other mounting position
Overpressure limit exceeded
Permissible temperatures
Correct the zero point via potentiometer
or key pad or software
Observe the permissible overpressure
limit
If complaint is unjustied, we will charge you the complaint processing fees.
8. Faults