Operating instructions/Betriebsanleitung, WUD-2x
7. Initial start-up and operation
Mechanical connection
Product label (example)
Coded manufacture
date
Use this spanner at for screwing in!
Don't use this spanner at!
P# Product No.
S# Serial No.
PIN assignment
Signal
Power Supply
GB
Remove the protection cap only just before installation.
When mounting the instrument, ensure that the sealing faces of the instrument and the measuring point are
clean and undamaged.
Do not scratch or nick the bead. Do not over tighten. Damage to the bead will aect the tting's performance
and may cause leakage in the system.
Screw in or unscrew the instrument only via the ats using a suitable tool and the prescribed torque. The
appropriate torque depends on the dimension of the pressure connection and on the sealing element used
(form/material). Do not use the case as working surface for screwing in or unscrewing the instrument.
When screwing the pressure transducer in, ensure that the threads are not jammed.
Mechanical Connection
Prepare the gas line connections appropriately.
You should blow clean all tting components (such as sealing gaskets, for example) using a clean/ltered gas. Please
refer to the specic technical guides furnished by the gasket manufacturers for additional specications.
You can then remove the protective lm, as well as any plastic caps there may be for protecting high-quality ttings.