14 WIKA operating instructions pressure transmitter model A-10
11218720.10 01/2012 GB/D/F/E
GB
3. Specications
Designation Ingress
protection
Wire
cross-section
Cable Ø Cable
material
Circular connector M12 x 1 (4-pin)
■
without mating connector IP 67 - - -
■
straight with solid laid cable IP 67 3 x 0.34 mm
2
4.4 mm PUR
■
angled with solid laid cable IP 67 3 x 0.34 mm
2
4.4 mm PUR
Cable outlet IP 67 3 x 0.34 mm
2
4 mm PUR
■
unshielded IP 67 3 x 0.34 mm
2
4 mm PUR
■
OEM version, unshielded IP 67 3 x 0.14 mm
2
2.85 mm TPU
The stated ingress protection (per IEC 60529) only applies when plugged in using mating connectors that have the appropriate ingress
protection.
Short-circuit resistance
S
+
vs. 0V
Reverse polarity protection
U
B
vs. 0V
Insulation voltage
DC 500 V
3.8 Materials
Wetted parts
Stainless steel 316L
from 10 bar 316L and 13-8 PH
Non-wetted parts
Stainless steel 316L, HNBR, PA66
For sealing materials see "Process connections"
Materials for electrical connections see "Electrical connections"