The conguration of the measuring cells diers depending on the
measuring range:
1
2
3
4
5
1
p
2
p
Fig. 7: Metallic measuring cell 10 mbar and 30 mbar - p
1
and p
2
process
pressures
1 Measuring element
2 Silicone diaphragm
3 Separating diaphragm
4 Filling oil
5 Integrated overvoltage arrester
1
p
2
p
1
2
3
4
Fig. 8: Metallic measuring cell from 100 mbar - p
1
and p
2
process pressures
1 Measuring element
2 Overload diaphragm/Middle diaphragm
3 Filling oil
4 Separating diaphragm
Power supply via the Probus DP/PA segment coupler. A two-wire
cable according to Probus specication serves as carrier of both
power and digital data transmission for multiple sensors. The
instrument prole of DPT10 corresponds to prole specication
version 3.0.
The GSD (instrument master les) and bitmap les necessary for
planning your Probus DP (PA) communication network are available
from the download section on the WIKA homepage
www.wika.com
under "Service". There you can also nd the appropriate certicates. In
Power supply and bus
communication
GSD/EDD
Dierential pressure transmitter DPT10 - metal measuring diaphragm Probus PA 11
3 Product description
37245-EN-100120