Wika UPT-21 Uživatelský manuál Strana 23

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 120
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 22
23
WIKA operating instructions process transmitter, model UPT-2x
14068347.01 06/2014 GB/D
GB
6 Commissioning, operation
Wire cross-section:
- Single wire: 0.13 ... 2.5 mm
2
- End splices: 0.13 ... 1.5 mm
2
- Ground screw, inside: 0.13 ... 2.5 mm²
- Ground screw, outside: 0.13 ... 4 mm²
- When using DIN 175301-803 A angular connector 1.5 mm² (only via end splices)
When using an M12 x 1 (4-pin) circular connector, the mating connector is supplied
by the customer. It is important to ensure the matching design from the connector
manufacturer.
6.2.2 Makingtheelectricalconnection
CAUTION!
Ingress of moisture.
Moisture can destroy the process transmitter.
- Protect the opened process transmitter against moisture.
1. Screw o the case head cover by hand and pull out the display and operating unit or
push-on cap.
2. Pass the connection cable through the cable gland and connect it.
- For pin assignment see chapter 6.2.3 “pin assignments”.
- Ground the process transmitter via the process connection or connect with the
equipotential bonding via the external grounding screw.
3. Tighten the cable gland.
- Recommended tightening torque 1.5 Nm
- Check that the seals are correctly seated in order to guarantee the ingress
protection.
- Ensure that no moisture can enter at the cable end.
Zobrazit stránku 22